Liquidmanagement AutoSampler
 
 
   
 
Contact: Mr. Chiu Tel:886-3-5783280 Ext.339
Product Description

The function of AD08-V200 Auto sampler is to feed specific amount (5~200μl) samples or reagents into the pits in the microwell plate accurately and rapidly to reduce the cycle time of biochemical testing.
AD08-200 is capable of providing sample pretreatment rapidly without cross-contamination and increasing reliability by reduction of human intervention.
AD08-V200 Sampler with user-friendly monitoring functions may be optimized to accommodate various testing sequences or kits via Mirle ELISA Program Editor. 
 

Specifications

‧ Dimensions: 930*640*730 mm
‧ Weight: 86 kg
‧ Control software: Mirle ELISA
‧ PC interface: RS232
‧ Power source: AC 100~240 V

 

Working Area

‧Sample rack: 96 sample tubes (5cc / 7cc tubes)
‧Reagent rack: 12 reagent troughs (12cc)
‧Microwell plate rack: 6 microwell plates (96 positions)
‧Pipette tip rack: 2 Pipette tip boxes (96 positions)
 

System Description

‧8 channels with variable distance for air-displacement sampling
‧Pipette head: ISO 8655-A-200 disposable pipette tip (200μl)
‧Pipette head working distance: (x, y, z) = (496, 354, 60)±0.1mm
‧Liquid Level Detection:
     Pressure type LLD (Liquid Level Detection)
     Measurable height of 7cc tube: 35mm(min)
     Measurable height of 5cc tube: 20mm(min)
‧Sample volume: 5~200μl
‧Sample feeding speed: Blood sample, under 7 minutes / microwell plate
  Reagent, under 2.5 minutes / microwell plate
‧Sample accuracy:

Sample volume Vs (μl) 5 10 50 100 200
Accuracy Es (%) ±10% ±7.5% ±2% ±1.5% ±1.0%
Precision CV (%) 7.50% 5% 1% 0.75% 0.75%